How the ABER biomass monitor works
Real time means real savings
Cells with intact plasma membranes in a fermenter can be considered to act as tiny capacitors under the influence of an electric field. The non-conducting nature of the plasma membrane allows a build up of charge. The resulting capacitance can be measured: it is dependent upon the cell type and is directly proportional to the membrane bound volume of these viable cells.
A unique range of probes
Choose from a range of in situ steam sterilisable probes manufactured in
robust materials that conform to FDA requirements & USP Class VI. The
smooth, annular ring electrode arrangement provides a surface free of areas of
entrapment and large EVEN electric field; ideal for many cell culture applications
where cells will collect in any well or crevice. For production applications ABER
offers a highly robust 25mm probe with flush electrodes- a design that is often
used in cGMP bioreactors for cell culture. ABER can also provide probes that can
be gamma irradiated for single use bioreactors. Option for a range of probes for
single use bioreactors including single use, sterilisable and flow through cell.
Specification
Frequency Range: | 50KHz to 20MHz |
Measuring Ranges: | Capacitance: 0.0 to 400pF/cm Conductivity: 1.0 to 40 mS/cm +/-0.1 mS/cm (Higher ranges available with compatible probes) |
Cell Concentration Range: | Depends on cell sizes but typically: Yeast (6 µm): 106 cells/ml to 1010 Cells/ml Bacteria (1 µm): 109 cells/ml to 1013 Cells/ml Animal Cell (12 µm): 105 cells/ml to 109 Cells/ml Plant Cell (50 µm): 103 cells/ml to 107 Cells/ml |
Resolution: | Resolution: 0.1 pF/cm. Approx: Bacteria typically 0.1 g/L dry weight or 2x109 Cells/ml for E. Coli. Yeast or Animal Cells 0.05g/L or 1 x105 Cells/ml The relationship of these capacitance values to biomass levels depends upon the cell type and line. |
Accuracy: | Typically bett er than ± 3% or ± 2% of the reading |
Stability: | Better than ± 0.2 pF/cm at constant temperature with standard conductivity solution of ~12 mS/cm |
Linearity: | Better than ± 1% over 100 pF/cm |
Precision: | Typically <±0.5 pF/cm, no filter active. |
Power Supply: | Natively 24V DC. This is typically supplied by an Aber Hub running on 110V AC to 240V AC mains. |
Environmental: | IP65 rated Recommended ambient operating temperature range: 5°C to 40°C |